Published in final edited form as: Mol Genet Metab. 2008 May 15;94(4):491–497. doi: 10.1016/j.ymgme.2008.03.019

Table 2.

Frataxin Lateral Flow Assay Intra- and Inter-assay Imprecision.

Intra-assay Imprecision
Experiment Dipstick 1 Dipstick 2 Dipstick 3 Mean SD CV (%)
Day 1 19,380 17,696 19,168 18,748 917 4.9
Day 2 19,867 19,535 18,260 19,221 848 4.4
Day 3 19,107 18,531 18,343 18,660 398 2.1
Day 4 21,502 23,164 23,717 22,794 1,152 5.1
Inter-assay Imprecision (Day 1- Day 4)
Mean Signal SD CV (%)
19,856 1,974 9.9

A single sample of normal cultured fibroblast whole cell extract (within working range of assay) was prepared as described in the Materials and Methods. Total protein concentration of the extract was determined by the BCA method, aliquots frozen at −80° C and then retrieved each day for analysis. Dipstick signal was determined using a Hamamatsu ICA immune-chormato reader and is expressed as Area units. Intra-assay imprecision was determined from n=3 dipsticks/ day. Inter-assay imprecision data was compiled from 4 separate days. Coefficient of variation (CV) was determined by the ratio of standard deviation (SD)/Mean × 100.