Skip to main content
. Author manuscript; available in PMC: 2009 Jul 1.
Published in final edited form as: Nat Protoc. 2009;4(4):582–591. doi: 10.1038/nprot.2009.28

Table 2.

Troubleshooting table

Step Problem Possible reason Solution
8 and 13;
Box 1,
step 4
and Box 2, step 7
Signal does not
increase with
respect to time or
cell number.
Signal reaches a
plateau
A high number of
cells are plated or a
large volume of
conditioned
medium is assayed
When performing the linearity of the reporter assay with
respect to time and cell number, the signal might reach a
plateau at the highest concentration of cells or at the later
time points (especially for SEAP). In this case, the
protocol needs to be scaled down either by assaying a
smaller volume of the conditioned medium and/or plating
lower amount of cells. However, it is not recommended to
assay less then 10 μl of conditioned medium as pipetting
errors may be higher especially with the Gluc assay
24 and
33
Dose-response
with respect to
drug
concentration is
not obtained
Too little or too
much drug is added
per well
Higher or lower doses of drug may be needed. In the case
of no response, use higher concentration of each
drug/well. In the case of high response even with the
lowest drug concentration, use less drug/well.
24;
Box 3,
step 6;
Box 4,
step 8
No reporter
activity is
observed in
blood/serum
Not enough cells
are implanted or
transduction
efficiency is low
More cells-expressing the reporter should be implanted. In
the case of tumor and other cells that are known to
proliferate in vivo, one may choose to wait 1 – 2 weeks to
obtain higher reporter signal before starting the drug
treatment.
33 No Gluc activity is
observed during
bioluminescence
imaging
Not enough cells
are concentrated in
one location of the
animal
Implant more cells. In the case of using cells which
proliferate in vivo, wait 2 weeks before the next imaging
session.
33 Signal obtained
from tumors is
saturated
Large tumor size Decrease the imaging time (10 s – 1min) in order to stay
within the linear range of bioluminescence imaging.
33,
Box 4,
step 8
In the case of
cells which are
known to
proliferate in vivo
(e.g., tumors)
SEAP activity
does not increase
over time
The number of cells
in vivo is above the
linear range of the
reporter assay
(especially for
SEAP). System
reached saturation
Dilute the serum sample in the assay buffer (20-fold
dilution rather then 4-fold) by mixing 2 μl of serum with 38
μl of assay buffer and proceed with the SEAP assay