Table 1.
Clinical Laboratory Reference Instruments at 35 U.S. Medical Centers
| Manufacturer (www.) | Reference instrument | Sites | Number of observations | Group | Bias (mg/dL) [SD, P], median (range) |
|---|---|---|---|---|---|
| Abbott (abbott.com) | Architect | 1 | 30 | − 1.47 | |
| Areoset | 1 | 25 | 55 | [10.09, 0.284] 1 (− 43 to +17) |
|
| Bayer (bayer.com) | Advia 1650 | 2 | 51 | 51 | − 3.82 [9.42, 0.006] − 3 (− 32 to + 18) |
| Beckman (beckman.com) | Beckman | 3 | 90 | ||
| LX20 | 5 | 169 | |||
| LXI | 1 | 37 | − 2.72 | ||
| DCx | 2 | 60 | 656 | [11.57, <0.001] | |
| Synchron | 2 | 70 | − 1 (− 96 to + 45) | ||
| LCX | 1 | 200 | |||
| CX-9 | 1 | 30 | |||
| Dade Behring (dadebehring.com) | RxL | 4 | 270 | − 0.64 | |
| Dimension | 1 | 30 | 384 | [7.48, 0.09] | |
| Xpand | 1 | 84 | − 1 (− 29 to + 43) | ||
| Johnson & Johnson (jnjgateway.com) | Vitros Fusion | 1 | 30 | ||
| Vitros | 4 | 215 | − 2.46 | ||
| Vitros 5.1 | 1 | 30 | 497 | [8.86, <0.001] | |
| Vitros 250 | 1 | 72 | − 2 (− 65 to + 16) | ||
| Vitros 950 | 1 | 150 | |||
| Olympus (olympusamerica.com) | AU series | 1 | 30 | 30 | − 7.03 [3.72, <0.001] −7 (−14 to 0) |
| Roche (rochediagnostics.com) | Integra | 1 | 30 | 30 | 0.567 [9.37, 0.743] 5 (− 28 to + 9) |
| Types of instruments | 20 | ||||
| Total medical centers | 35 | ||||
| Total multicenter observations | 1,703 | ||||
| Mean bias (mg/dL), all observations [SD, P] | − 2.2 [9.8, <0.001] | ||||
| Median bias (mg/dL), all observations [range] | − 1 [− 96 to + 45] |