Table 1.

Clinical Laboratory Reference Instruments at 35 U.S. Medical Centers

Manufacturer (www.) Reference instrument Sites Number of observations Group Bias (mg/dL) [SD, P], median (range)
Abbott (abbott.com) Architect 1 30   − 1.47
  Areoset 1 25 55 [10.09, 0.284]
1 (− 43 to +17)
Bayer (bayer.com) Advia 1650 2 51 51 − 3.82
[9.42, 0.006]
− 3 (− 32 to + 18)
Beckman (beckman.com) Beckman 3 90    
  LX20 5 169    
  LXI 1 37   − 2.72
  DCx 2 60 656 [11.57, <0.001]
  Synchron 2 70   − 1 (− 96 to + 45)
  LCX 1 200    
  CX-9 1 30    
Dade Behring (dadebehring.com) RxL 4 270   − 0.64
  Dimension 1 30 384 [7.48, 0.09]
  Xpand 1 84   − 1 (− 29 to + 43)
Johnson & Johnson (jnjgateway.com) Vitros Fusion 1 30    
  Vitros 4 215   − 2.46
  Vitros 5.1 1 30 497 [8.86, <0.001]
  Vitros 250 1 72   − 2 (− 65 to + 16)
  Vitros 950 1 150    
Olympus (olympusamerica.com) AU series 1 30 30 − 7.03
[3.72, <0.001]
−7 (−14 to 0)
Roche (rochediagnostics.com) Integra 1 30 30 0.567
[9.37, 0.743]
5 (− 28 to + 9)
 Types of instruments 20        
 Total medical centers   35      
 Total multicenter observations     1,703    
 Mean bias (mg/dL), all observations [SD, P]         − 2.2 [9.8, <0.001]
 Median bias (mg/dL), all observations [range]         − 1 [− 96 to + 45]