Table 3.

Stability of the 250 GHz operating parameters during CW operation

Technical parameter Average value Standard deviation (%)
Filament voltage (V) 3.274 0.004 0.1

Filament current (A) 4.515 0.008 0.2
Beam voltage (kV) 12.090 0.047 0.4
Pressure (10−9 Torr) 3.430 0.019 0.5
Frequency (GHz) 250.559348 0.000359 0.000143
Power (diode units (mV)) 229.75 1.94 0.8

Note from Table 3 that the power was stable to ~0.8% and the other parameters were well below this level including the frequency which showed a stability of 1.6 ppm.