Skip to main content
. 2009 May 8;75(13):4297–4306. doi: 10.1128/AEM.02549-08

TABLE 1.

Estimated LOD by surface material and sampling method

Surface Method No. of runs No. of samples LODa
CFU/100 cm2
CFU/sampled areab
LOD95 95% CI LOD95 95% CI
Steel Swabc 9 107 190 74-3,700 200 76-3,800
Wiped 10 90 15 7.6-84 140 71-780
Vacuumd 10 90 44 24-140 410 230-1,300
Carpet Swab 9 108 40 16-560 41 17-580
Wipe 10 (9)e 90 (81) 9.2 (9.9) 1.8-2.7 × 1031f (4.6-760) 85 (92) 17-2.5 × 1032f (42-7,100)
Vacuum 10 90 28 14-130 260 130-1,300
a

The LOD, defined here as the lowest concentration that could be detected (i.e., positive for CFU) with LOD95, was estimated using probit regression (SAS PROBIT procedure), and 95% CIs were estimated using Fieller's procedure (15).

b

The sampled surface areas were approximately 103 cm2 for the swab method and 929 cm2 for the wipe and vacuum methods.

c

Steel-swab combination results exclude a single sample with contaminant overgrowth.

d

The steel-wipe and steel-vacuum combinations required 3 or more CFU for a positive result (see text for details); all other surface-method combinations required 1 or more CFU for a positive result.

e

The carpet-wipe combination CI was very wide, likely due to the results of a single run in which 100% of the wipe samples were positive in spite of the very low surface loading; results excluding this run are in parentheses.

f

The 95% CI for the carpet-wipe combination could not be computed due to constraints imposed by Fieller's method; results shown have 94% confidence rather than 95% confidence.