. 2009 Jun;64(6):585–590. doi: 10.1590/S1807-59322009000600015

Table 1 -.

Ventilator settings employed and the corresponding percentage recovery of absolute humidity (%RAH) for each HME

VT (mL) Flow-rate (L/min) f (/min) FiO2 %RAH
G2S Portex HCH Pall G light G compact Hygrobac S Hygroster
200 30 10 0.5 94.4 95.5 95.2 95.1 94.1 94.7 96.4 94.2
500 30 10 0.5 88.3 85.5 88.9 88.3 87.9 89.7 92.3 85.8
1000 30 10 0.5 80.5 74.9 76.4 68.8 79.9 81.7 85.4 74.5

500 30 5 0.5 87.8 85.6 86.3 83.6 86.4 89.0 91.0 83.4
500 30 10 0.5 88.5 85.8 87.6 84.8 87.7 90.4 92.0 85.0
500 30 20 0.5 90.2 85.9 89.6 87.3 88.2 91.5 92.8 87.1

500 30 10 0.5 88.0 85.9 87.4 85.0 87.8 90.3 92.2 87.1
500 60 10 0.5 89.1 87.1 89.2 87.5 88.5 91.6 92.9 90.2
500 90 10 0.5 90.5 87.4 90.1 90.8 90.0 92.5 93.7 92.0

Note: VT = tidal volume; f = respiratory rate; FiO2 = fraction of inspired oxygen; %RAH = percentage recovery of absolute humidity