. 2009 May 6;47(7):2194–2199. doi: 10.1128/JCM.01264-08

TABLE 2.

Experimental results for repetitively quantified (20 times) serial dilutions of HIV-1 DNA standards in LightCycler 1.0 and LightCycler 2.0 platformsa

Exptl platform No. of HIV-1 DNA copies per reaction No. of HIV-1 DNA copies/106 PBMCs Mean ± SD of exptl valuesb CV (%) of exptl valuesb
LightCycler 1.0 106 4 × 106 5.99 ± 0.08 1.36
105 4 × 105 5.08 ± 0.20 3.86
104 4 × 104 3.95 ± 0.09 2.16
103 4 × 103 2.99 ± 0.15 5.02
102 400 2.21 ± 0.27 12.37
25 100 1.44 ± 0.27 18.87
12.5 50 0.79 ± 0.21 25.93
LightCycler 2.0 10 40 1.03 ± 0.13 12.99
6.25 25 0.87 ± 0.31 35.32
3.125 12.5 0.17 ± 0.40 232.26
a

Each reaction was performed in a multiplex setting with 5 × 105 copies of the CCR5 gene in a final reaction volume of 25 μl. The experimental results were 100% positive for all concentrations with both platforms.

b

Log10-transformed values.