Table 3.

Correlation Sample RealTime Results Summary

Assay input volume
1 ml 0.5 ml 0.6 ml
Tested 171 135 180
Delayed IC cycle number 14 2 2
Repeated 3 0 0
Detected but not quantified 19 25 13
Not detected 8 10 10
Total for Deming regression 133 98 155

The results of testing samples previously assayed using Amplicor are shown. One hundred seventy-one (171), 135, and 180 samples were tested in the 1.0 ml, 0.5 ml, and 0.6 ml RealTime assays, respectively. Samples with delayed IC cycle numbers produced errors that prevented results from being calculated; a small number of samples had sufficient volume to permit repeat testing. Samples that were “Detected but not quantified” or “Not detected” were not retested due to insufficient volume.