Skip to main content
. Author manuscript; available in PMC: 2009 Jul 16.
Published in final edited form as: Nat Protoc. 2007;2(3):512–522. doi: 10.1038/nprot.2007.5

Table 1.

Troubleshooting table.

Problem Possible reason Solution
No signal Defective secondary Ab Follow manufacturer’s instructions to rehydrate the secondary Ab. Mix 10 µl of a
1:10,000 dilution of the secondary Ab in washing buffer and 50 µl of the developing
reagent in a microplate, and read the chemiluminescence
Not enough DNA loaded in
each well
Run positive controls and load different concentrations of the purified DNA nitrone
adducts
No DNA nitrone adducts Run a positive and a negative control in each plate. Do not incubate the nitrone
adducts with Reacti-Bind DNA coating solution for >6 h
Treatments bleach DNA nitrone
adducts*
Extract the positive control under the same conditions used to extract DNA from
samples
High background Excessive amount of DNA
loaded in each well
Use 25 µl of a 5–10 µg µl−1 DNA solution. Use the type of development reagent
suggested in this protocol
If purified DNA from blood or
tissue, beware of hemopro-
teins as contaminants
Add 10 µl of DNA extracted from cells or tissues without any treatment and 50 µl of the
development reagent. If hemoproteins are present, there will be a chemiluminescent
signal. Re-purify DNA sample
Poor reproducibility Operational failures Use multichannel micropipettes and an automatic washer. Include a positive control
sample
*

The term bleaching refers here to an alteration of the epitope (the nitrone motif) by unknown chemical mechanisms with a consequent loss of Ab binding.