Table 1.
Troubleshooting table.
| Problem | Possible reason | Solution |
|---|---|---|
| No signal | Defective secondary Ab | Follow manufacturer’s instructions to rehydrate the secondary Ab. Mix 10 µl of a 1:10,000 dilution of the secondary Ab in washing buffer and 50 µl of the developing reagent in a microplate, and read the chemiluminescence |
| Not enough DNA loaded in each well |
Run positive controls and load different concentrations of the purified DNA nitrone adducts |
|
| No DNA nitrone adducts | Run a positive and a negative control in each plate. Do not incubate the nitrone adducts with Reacti-Bind DNA coating solution for >6 h |
|
| Treatments bleach DNA nitrone adducts* |
Extract the positive control under the same conditions used to extract DNA from samples |
|
| High background | Excessive amount of DNA loaded in each well |
Use 25 µl of a 5–10 µg µl−1 DNA solution. Use the type of development reagent suggested in this protocol |
| If purified DNA from blood or tissue, beware of hemopro- teins as contaminants |
Add 10 µl of DNA extracted from cells or tissues without any treatment and 50 µl of the development reagent. If hemoproteins are present, there will be a chemiluminescent signal. Re-purify DNA sample |
|
| Poor reproducibility | Operational failures | Use multichannel micropipettes and an automatic washer. Include a positive control sample |
The term bleaching refers here to an alteration of the epitope (the nitrone motif) by unknown chemical mechanisms with a consequent loss of Ab binding.