Abstract
Background
Fluid handling technology is acquiring an ever more prominent place in laboratory science whether it is in simple buffer exchange systems, perfusion chambers, or advanced microfluidic devices. Many of these applications remain the providence of laboratories at large institutions with a great deal of expertise and specialized equipment. Even with the expansion of these techniques, limitations remain that frequently prevent the coupling of controlled fluid flow with other technologies, such as coupling microfluidics and high-resolution position and force measurements by optical trapping microscopy.
Method
Here we present a method for fabrication of multiple-input laminar flow devices that are optically clear [glass] on each face, chemically inert, reusable, inexpensive, and can be fabricated on the benchtop in approximately one hour. Further these devices are designed to allow flow regulation by a simple gravity method thus requiring no specialized equipment to drive flow. Here we use these devices to perform total internal reflection fluorescence microscopy measurements as well as position sensitive optical trapping experiments.
Significance
Flow chamber technology needs to be more accessible to the general scientific community. The method presented here is versatile and robust. These devices use standard slides and coverslips making them compatible with nearly all types and models of light microscopes. These devices meet the needs of groups doing advanced optical trapping experiments, but could also be adapted by nearly any lab that has a function for solution flow coupled with microscopy.
Introduction
Coupling of optical trapping with flow chambers, lab-on-a-chip, and other microfluidic devices has been accomplished by a number of groups[1]–[6] for two major purposes; cell sorting and manipulation[1]–[4] and making biochemical and biophysical measurements[5], [6]. Most of the cell sorting devices use traditional fabrication methods and have one optically uniform glass face while the channel walls and opposing face are made of a polymer such as polydimethylsiloxane (PDMS)[7].
Traditional microfluidic fabrication methods have several limitations. First, the fabrication method requires specialized equipment not readily available to many groups. Second, the PDMS face opposite the coverslip prevents some optical techniques that require accurate visualization through both sides of the device. One such technique is optical trapping with precision force and position measurements, because the PDMS layers are of non-uniform density and distort the wavefront of the required detection lasers in unpredictable ways. This can make experiments such as the study of molecular motor stepping[8], [9] or forced unfolding[10] problematic in such devices. Since buffer conditions can change these behaviors, being able to quickly change the buffer conditions while studying the same molecules could be useful but remains illusive.
As a result of these difficulties most trapping experiments that require precision measurements are performed using chambers made of a slide and coverslip linked by a piece of double sided tape and sealed with vacuum grease[11]. When multiple conditions are required, multiple independent experiments are run. While very effective these devices have limitations as well. Experiments using sealed sticky tape devices require premixing all components, so unwanted component self-association prior to visualization is a problem. It is also difficult to make complex geometry devices using tape flow cells. Further, coupling tape devices to flow often requires drilling through glass and attaching ports, which can interfere with optical components such as oil condensers[12].
For biochemical and biophysical measurements with flow other devices have been produced. The Kowalczykowski group has coupled optical trapping and flow technology to generate a platform for performing fluorescence microscopy measurements on proteins bound to a single DNA molecule[5]. However these devices are made from etched glass, which requires special facilities to fabricate and have not been, to our knowledge, coupled to high-resolution detection systems.
Recently the Wuite group developed a system to couple flow and optical trapping with high-resolution position and force detection[6] to examine the single molecule behavior of DNA binding proteins. They report using parafilm to construct channels. In our hands uniform parafilm adherence was problematic so these devices tended to leak. We also had difficulties incorporating fluid ports that would not interfere with our optics. Our research requires a device of this type that prevents reagent mixing and self-assembly but allows trapped particles to be transported into different environments. Here we present a simple manual method for flow chamber fabrication that can be preformed on a benchtop and is fast, inexpensive, requires no glass drilling, produces devices that are reusable and meet the requirements of advanced optical trapping assays, including all-glass optical faces, and a radial arrangement of tubing connections that allows us to use oil-immersion condensers and objectives.
Methods
Device Fabrication
Our flow chamber design uses a single flat sheet of silicon rubber, which is cut by hand into the desired shape. The advantage of using commercially available silicon rubber sheets is that they have a uniform thickness. Moreover, they are easily cut by hand and manipulated, since they only adhere once activated by oxygen plasma. We cut out a piece of silicon rubber (McMaster-Carr, 87315K63) of the same dimensions as the coverslips that will be used (25 mm round) (see Supplemental Material S1 for more detailed illustrated step-by-step assembly and usage protocols). Next the desired pattern is cut into a silicon rubber sheet using sharp razors (Pattern in Figure 1a). Channel entrances should be left intact at this point so that sheet remains in one piece until after it is joined to the slide (Figure 1). The cut sheet constitutes the device body.
The microscope slide and the device body should be rinsed with pure, deionized water and dried to remove large debris. Plasma clean the slide and device body to functionalize the surfaces (acid washing will also work eliminating the need for a plasma cleaner[13]). Once cleaned press the device body onto the slide. The bond is permanent. Cut channel entrances open with the razor.
Next plasma clean the device complex and coverslip and press them together. This produces a sandwich of rubber with channel openings between the slide and coverslip. Thin wall PEEK tubing (Upchurch, 1569) is then connected to the device. Compress tubing at one end and slide into channel entrances. Seal the joint between the device and the tubing with quick drying epoxy. Once dry the device is complete and can be hooked up to a flow driving system via HPLC adaptors. All devices should be tested for leaks and correct flow characteristics before use in experiments.
Device Testing
Since the devices are not precision cut or cast, there will be variance from device to device. Thus, if very specific or complex flow characteristics are required this technique may not be suitable. Additionally, each flow cell should be characterized individually. For example, in multiple input laminar flow cells, if there is variance in the structure of the input channels or the length of the tubing leading from the reservoir the width and flow rate of each lane in the main channel may vary slightly. Even with these limitations for many applications this method is sufficient.
Flow rates can be measured in several ways. For high flow rates [>1 ul/sec] driving a set amount of fluid through the device and measuring the time required to pass a given volume is an effect method. For extremely slow flow rates, observing the motion of particles such as fluorescent beads as they transit a known distance as visualized by the microscope can give an accurate estimation of flow rate. It is important to note that flow rates are fastest in the center of the channels and slower near surfaces. If driving flow by gravity it is important to remember that rates will drop as the column heights balance. Once at a desired flow rate, the water column height must be routinely adjusted to maintain the flow rate.
Device Cleaning
Since these systems are entirely composed of PEEK, silicon rubber and glass they are very chemically inert. As such they can be cleaned very rigorously allowing them to be reused without fear of contamination. We rinse the systems sequentially with water, 1 M NaOH, EtOH then finally water again (see Supplemental Material S1 for details). This cleans all internal surfaces.
Myosin Motility Assay
This assay was performed in two ways: in a flow chamber and a series of sealed chambers (the current standard method). Data in panels 3b and 3c were collected on different days but used the same motor preparation and solution stocks. Incubation times and all other variables were kept as similar as possible.
For the flow chamber experiments, a four input laminar flow chamber was used. The device was primed with water by injecting it into the device via the output port using a syringe. All setup solutions are added in this way. This eliminates bubbles and insures all inputs are exposed to the same conditions. Anti-GFP antibody (QBiogene, 50 ng/µl) was added into the device to coat all of the internal surfaces[14]. BSA was then added as a blocking agent (1 mg/ml). GFP tagged Myosin VI heavy meromyosin (HMM) dimer [14] was added followed by filamentous actin (chicken skeletal muscle, 100 nM [15]), and then assay buffer (25 mM imidazole, pH 7.5, 25 mM KCl, 1 mM EGTA, 4 mM MgCl2, 10 mM DTT, 0.86 mg/ml glucose oxidase, 0.14 mg/ml catalase, 9 mg/ml glucose) with zero ATP. The microscope was focused and a desirable field of view was found. Assay buffers with desired ATP concentrations (0, 0.1, 1, 2 µM) were loaded into the input reservoirs. Movies were then recorded for the zero ATP condition. The valve to the 0.1 µM ATP containing solution was then opened, a sufficient replacement volume was allowed to flow through and a movie was recorded. Data in panel 3b were recorded with the flow turned off. Data in panel 3c were recorded with flow off then with flow on. This process was repeated for each ATP solution. Data was recorded on a home built total internal reflection fluorescence (TIRF) microscope with Andor iXon camera. Movies were analyzed using ImageJ.
Optical Trapping Assay
All experiments were performed on a home built fluorescence and optical trapping microscope.
Power Spectra
A two input device was used. The system was primed as described above. 1 µm beads were loaded into one input reservoir, and assay buffer into the other. Beads were flowed into the device and a single bead was trapped. Flow rate was approximately 1 µL/sec. The trapped bead was moved into the buffer channel and the bead position was recorded at 10 kHz, without a lowpass antialias filter. Flow was then turned off and the power spectrum was repeated with the same bead. Power spectra were calculated using the Igor Pro software package.
Dumbbell
A four input device was used. BSA blocking solution was pumped into the output port of the device as described above to prime the device and block the surface. 1 µm neutravidin coated beads (biotinylated beads from Invitrogen incubated in a neutravidin solution and blocked with BSA) were loaded into one input reservoir, TMR-phalloidin (Sigma-Aldrich) stabilized actin filaments with ten percent of the monomers labeled with biotin were loaded into the second input reservoir, and buffer was loaded into the third and fourth. The solutions were allowed to flow into the device. Beads were trapped then the stage was moved so that those beads transited into the actin containing channel. Once an actin filament was captured (by flowing against a bead) the stage was again moved so the beads and captured filament moved into the buffer channel where the dumbbell was assembled. The bead with the attached filament was placed upstream. Buffer flow extended the filament. The second bead was then moved into position at the free end of the extended filament and the bond was formed.
Results
Here we present data from several experiments that show the broad applicability and high degree of functionality of these devices. This data shows that these devices are functional for a broad range of microscopy experiments including sensitive optical trapping experiments.
Device Characteristics
Devices were fabricated using main channel widths of 0.3 up to 2 mm, input channel widths of 0.5 to 1.0 mm, and between 1 and 5 inputs. All devices produced laminar flow, where each solution in the main channel remained separate and only mixed via diffusion (Figure 2). Devices were tested over a range of flow rates from greater than a 1 milliliter per second to rate of below 1 microliter per minute.
Myosin Motility Assay with Multiple ATP Concentrations
As a proof-of-principle experiment these devices were used to perform myosin gliding filament assays under varying ATP concentrations using TIRF microscopy. Myosin VI (as is true of all myosins) shows an ATP concentration dependent motility rate [16], [17]. The device was coated with myosin VI before actin filaments were added. Some filaments were bound by the motors at the coverslip surface and immobilized in the absence of ATP. A series of motility buffers with increasing ATP concentrations (0, 0.1, 1, 2 µM) was then allowed to flow into the flow chamber and movies were recorded at each ATP concentration, with and without flow. The addition of ATP caused the motors to move (Figure 3). The motility profiles were similar between the two assays.
There are several advantages to the flow chamber method. This method allowed us to assess the ATP-dependent motility of a single population of motors, rather than preparing a different slide to analyze each ATP concentration. This prevents slide-to-slide variation, making results easy and reliable to interpret. Open chamber devices (sticky tape) can dry out during the course of an experiment as happens if inadequately sealed, changing ATP concentration. The flow chambers presented here do not share this limitation. Finally, continuous solution flow can also be used to prevent the accumulation of ADP in the chamber, which could lower the apparent motility rate of the motor. The shear induced by flow does not appear to affect myosin VI motility but it should be considered if assaying other motors in the presence of flow.
Optical Trapping Assays
To demonstrate that optical trapping, manipulation, and position detection is practical with these devices two assays were performed. 1) We trapped 1 µm polystyrene beads and measured power spectrums with and without flow. 2) We assembled actin dumbbells and manipulated them. The first experiment uses optical trapping, bright field microscopy and laser position detection. The second uses optical trapping, bright field and epifluorescent microscopy.
Power Spectra
A bead was trapped under flow of 1 µL/sec flow rate. A power spectrum was recorded [8], [18], [19]. Then flow was stopped and another power spectrum was measured. The power spectra are nearly indistinguishable in the X direction (the direction of flow) (Figure 4). First this demonstrates that trapping is indeed possible in these devices. Second these devices are functional for position detection and force measurements when the system is under flow and when flow is stopped. This creates a dynamic environment in which to design and perform experiments.
Dumbbells
For many actin based optical trapping experiments, such as analysis of myosin motors stepping, forming an actin dumbbell is required [20], [21]. A dumbbell is an actin filament with each end connected to beads held in separate optical traps. Here we trapped two beads under flow, attached an actin filament to one of those beads under flow, then moved the second bead into position downstream behind the trapped filament to form the dumbbell (Figure 5, Supplemental Movie S1). The trapped filament can be manipulated and used for desired assays.
Unlike traditional methods where beads and actin filaments are mixed together in extremely low dilutions and the user must find and assemble them (a process which can be time sensitive), using the flow based method present here each component is added in isolation and can be assembled extremely quickly (as fast as 13 seconds for a dumbbell in our hands). Further, since a buffer channel is present which lacks any other components, these structures can be manipulated without concern for accidental addition of unwanted components (another bead or actin filament sticking to the trapped dumbbell for example). This ability increases the percentage of experiments that are successful under most circumstances. Additional flow lanes can be added to allow for addition of other components to the trapped scaffold, as was demonstrated by the Wuite group's DNA and H-NS experiments[6].
Discussion
Design and Potential Applications
Beyond the ability to perform specialized optical trapping assays these devices have other important virtues. The fabrication technique reported here requires no glass drilling, no training, and little to no specialized equipment (plasma cleaner is recommended). Devices are constructed using standard microscope slides and coverslips making them compatible with nearly all light microscopes. The assembly protocol is general enough that any glass coverslip can be accommodated, allowing for matching of coverslips to objectives for optimum results and making design of devices of different dimensions simple. The slide can even be replaced by a second coverslip to accommodate dual objective optical trapping microscopes. The fluid volume of the devices is also flexible ranging from 30 down to 3 microliters depending on design (see Supplemental Material S1 for ways to lower device volume). After most usages devices can be cleaned and reused. Because of the uniform nature of the silicone rubber sheets used to make these devices, when the devices are properly assembled the coverslips lay extremely flat allowing for minimal focal drift as the sample is moved. All fluorescence imaging characteristics seem to closely mimic those of sealed sticky tape flow cells, including low oxygen permeability yielding long fluorescence lifetimes and the ability to clearly resolve single molecules.
These devices do carry inherent limitations, which should not be overlooked. Since channels are cut by hand variation from device to device will occur. For some applications this will require that each device be tested before use. Complex multilayer applications such as those requiring pneumatic microvalves [22], [23] cannot be adapted to work with these devices. Finally device volumes are typically around 10 microliters and require a reservoir to generate flow, so applications that use very small volumes are not practical.
We see several areas in which these devices could be immediately applicable. The flexibility in device design, low cost, and the ease of construction should allow them to be adapted by groups in many fields. Laminar flow chambers, from single to many inputs, can be made exactly as described here. Coupling valves to the fluid handling system allows for many possibilities from side-by-side laminar flow to iterative additions of different solutions.
Coupled with valves these devices allow for rapid and repeated changes in buffer conditions. With computer-controlled valves, experiments or condition screens could be automated saving researchers time and effort. Our system uses commercially available solenoid valves (Lee Company) to control inputs.
Most current optical trapping experiments use sealed chambers. This necessitates mixing all experimental components together before starting the experiment. If these components interact with each other in solution then it becomes difficult to control and study these interactions. Devices like these allow for iterative assembly of complex systems, opening up new avenues of exploration for single molecule scientists[6]. As our knowledge increases study of more complex systems becomes necessary. These devices should serve as a useful tool for scientist attempting to do that.
Supporting Information
Acknowledgments
The authors would like to thank Crista Brawley for preparing the myosin VI used in the motility assays. DSC would like to thank Wesley Jun and Stas Nagy for the use of their camera equipment.
Footnotes
Competing Interests: The authors have declared that no competing interests exist.
Funding: This work was funded by an American Heart Association Fellowship (to D.S.C.), a Burroughs-Wellcome Career Award at the Scientific Interface (R.S.R.), and a grant from the NIH (R.S.R., GM078450). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1.Enger J, Goksor M, Ramser K, Hagberg P, Hanstorp D. Optical tweezers applied to a microfluidic system. Lab Chip. 2004;4:196–200. doi: 10.1039/b307960k. [DOI] [PubMed] [Google Scholar]
- 2.Perroud TD, Kaiser JN, Sy JC, Lane TW, Branda CS, et al. Microfluidic-based cell sorting of Francisella tularensis infected macrophages using optical forces. Anal Chem. 2008;80:6365–72. doi: 10.1021/ac8007779. [DOI] [PubMed] [Google Scholar]
- 3.Murata M, Okamoto Y, Park YS, Kaji N, Tokeshi M, et al. Cell separation by the combination of microfluidics and optical trapping force on a microchip. Anal Bioanal Chem. 2009 doi: 10.1007/s00216-009-2648-5. [DOI] [PubMed] [Google Scholar]
- 4.Roman GT, Chen Y, Viberg P, Culbertson AH, Culbertson CT. Single-cell manipulation and analysis using microfluidic devices. Anal Bioanal Chem. 2007;387:9–12. doi: 10.1007/s00216-006-0670-4. [DOI] [PubMed] [Google Scholar]
- 5.Galletto R, Amitani I, Baskin RJ, Kowalczykowski SC. Direct observation of individual RecA filaments assembling on single DNA molecules. Nature. 2006;443:875–8. doi: 10.1038/nature05197. [DOI] [PubMed] [Google Scholar]
- 6.Dame RT, Noom MC, Wuite GJ. Bacterial chromatin organization by H-NS protein unravelled using dual DNA manipulation. Nature. 2006;444:387–90. doi: 10.1038/nature05283. [DOI] [PubMed] [Google Scholar]
- 7.McDonald JC, Duffy DC, Anderson JR, Chiu DT, Wu H, et al. Fabrication of microfluidic systems in poly(dimethylsiloxane). Electrophoresis. 2000;21:27–40. doi: 10.1002/(SICI)1522-2683(20000101)21:1<27::AID-ELPS27>3.0.CO;2-C. [DOI] [PubMed] [Google Scholar]
- 8.Neuman KC, Block SM. Optical trapping. Rev Sci Instrum. 2004;75:2787–809. doi: 10.1063/1.1785844. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Visscher K, Schnitzer MJ, Block SM. Single kinesin molecules studied with a molecular force clamp. Nature. 1999;400:184–9. doi: 10.1038/22146. [DOI] [PubMed] [Google Scholar]
- 10.Kellermayer MS, Smith SB, Granzier HL, Bustamante C. Folding-unfolding transitions in single titin molecules characterized with laser tweezers. Science. 1997;276:1112–6. doi: 10.1126/science.276.5315.1112. [DOI] [PubMed] [Google Scholar]
- 11.Kron SJ, Spudich JA. Fluorescent actin filaments move on myosin fixed to a glass surface. Proc Natl Acad Sci U S A. 1986;83:6272–6. doi: 10.1073/pnas.83.17.6272. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Brewer LR, Bianco PR. Laminar flow cells for single-molecule studies of DNA-protein interactions. Nature Methods. 2008;5:517–25. doi: 10.1038/nmeth.1217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Campbell K, Groisman A. Generation of complex concentration profiles in microchannels in a logarithmically small number of steps. Lab Chip. 2007;7:264–72. doi: 10.1039/b610011b. [DOI] [PubMed] [Google Scholar]
- 14.Rock RS, Ramamurthy B, Dunn AR, Beccafico S, Rami BR, et al. A flexible domain is essential for the large step size and processivity of myosin VI. Mol Cell. 2005;17:603–9. doi: 10.1016/j.molcel.2005.01.015. [DOI] [PubMed] [Google Scholar]
- 15.Pardee JD, Spudich JA. Purification of muscle actin. Meth Enzymol. 1982;85:164–81. doi: 10.1016/0076-6879(82)85020-9. [DOI] [PubMed] [Google Scholar]
- 16.Rock RS, Rice SE, Wells AL, Purcell TJ, Spudich JA, et al. Myosin VI is a processive motor with a large step size. Proc Natl Acad Sci U S A. 2001;98:13655–9. doi: 10.1073/pnas.191512398. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Sellers JR. Myosins: a diverse superfamily. Biochim Biophys Acta. 2000;1496:3–22. doi: 10.1016/s0167-4889(00)00005-7. [DOI] [PubMed] [Google Scholar]
- 18.Svoboda K, Schmidt CF, Schnapp BJ, Block SM. Direct observation of kinesin stepping by optical trapping interferometry. Nature. 1993;365:721–7. doi: 10.1038/365721a0. [DOI] [PubMed] [Google Scholar]
- 19.Mehta AD, Finer JT, Spudich JA. Detection of single-molecule interactions using correlated thermal diffusion. Proc Natl Acad Sci U S A. 1997;94:7927–31. doi: 10.1073/pnas.94.15.7927. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Finer JT, Simmons RM, Spudich JA. Single myosin molecule mechanics: piconewton forces and nanometre steps. Nature. 1994;368:113–9. doi: 10.1038/368113a0. [DOI] [PubMed] [Google Scholar]
- 21.Rock RS, Rief M, Mehta AD, Spudich JA. In vitro assays of processive myosin motors. Methods. 2000;22:373–81. doi: 10.1006/meth.2000.1089. [DOI] [PubMed] [Google Scholar]
- 22.Studer V, Hang G, Pandolfi A, Ortiz M, Anderson WF, et al. Scaling properties of a low-actuation pressure microfluidic valve. Journal of Applied Physics. 2004;95:393–398. [Google Scholar]
- 23.Oh KW, Ahn CH. A review of microvalves. J Micromech Microeng. 2006;16:R13–R39. [Google Scholar]