. 2009 May 19;13(3):R73. doi: 10.1186/cc7884

Table 2.

Comparison between the three methods restricted to the very stable period

Maximum Minimum Mean SD
CO values
 PAC-CCO 8.7 2.5 4.8 1.4
 NICOM™ 10.2 2.1 4.8 1.4
 FloTrac-Vigileo™ 11.2 1.0 5.0* 1.2
Differences in CO
 Vigileo™ – PAC-CCO 4.8 -3.0 0.1 0.9
 NICOM™ – PAC-CCO 5.3 -2.7 0.0 1.0
 NICOM™ – Vigileo™ 4.1 -5.1 -0.1 1.2
Relative errors
 Vigileo™ 0.97 0.0 0.16 0.12
 NICOM™ 1.15 0.0 0.17 0.12

Relative error = (tested CO – PAC-CCO)/PAC-CCO.

* P < 0.05 vs. PAC-CCO, P < 0.05 for NICOM™ vs. Vigileo™

CO = cardiac output; SD = standard deviation.