Published in final edited form as: J Biomed Opt. 2008;13(5):054031. doi: 10.1117/1.2976432

Table 1.

Comparison of reconstructed hemoglobin amplitudes. The maximum (minimum) response amplitude was calculated for a region of interest in each of the DOT alone and fusion reconstructions for the five subjects. Regions of interest were manually defined in the motor cortex area. The BOLD calibration factor (α) was calculated as described in the text. The values shown are for the reconstructions at a regularization 1/Γ = 10 µM.

DOT-only Data fusion BOLD calibration (α)
Subject ΔHbO2 ΔHbR ΔHbO2 ΔHbR
A 5.5 µM −2.9µM 6.1µM −1.2µM −0.41 %-BOLD/µM
B 0.3 µM −1.8µM 3.9µM −1.5µM −0.52 %-BOLD/µM
C 3.1 µM −1.4µM 4.4µM −0.7µM −1.24 %-BOLD/µM
D 4.9 µM −1.0µM 4.5µM −0.3µM −0.19 %-BOLD/µM
E 7.3 µM −3.µM 1.6µM −0.5µM −0.39 %-BOLD/µM
Group 4.2 µM −2.0µM 4.1µM −0.9µM −0.55 %-BOLD/µM