Table 1.
Preset name | Atlasa | Grid | Square | Hole |
---|---|---|---|---|
Example image | ||||
Magnification | - | 100x | 480x | 3kx |
Size (pixels/side) | - | 512 | 512 | 512 |
Defocus (μm) | 0 | 0 | 0 | -100 |
Pixel size after binning (nm) |
- | 840 | 170 | 28 |
Dose delivered to specimen (e-/Å2) |
negligible | negligible | negligible | <0.01 |
Field of view (μm/side) |
∼2000 | 430 | 90 | 14 |
Description/Use | Selecting viable grid squares | Building the atlas | Selecting regions with good holes |
Selecting potential targets and corresponding focus positions |
Preset name | Tomography Preview | Focus | Tomography | |
---|---|---|---|---|
Example image | ||||
Magnification | 22.5kx | 27.5kx | 27.5x | |
Size (pixels/side) | 1024 | 1024 | 2048 | |
Defocus (μm) | -20 | - | -10 | |
Pixel size after binning (nm) |
1.9 | 1.6 | 0.78 | |
Dose delivered to specimen (e-/Å2) |
<0.25 | 0b | 1-3c | |
Field of view (μm/side) |
2.0 | 1.6 | 1.6 | |
Description/Use | Highly defocused to enhance contrast / Final selection and prioritization of targets |
Adjusting defocus and stage z-height |
Acquiring the tilt series |
Atlas is not technically a preset but actually a montage of Grid images.
Focus targets are selected to be out of the specimen area.
Tomography preset dose depends on the tilt series dose and current tilt angle.