Skip to main content
. 2009 Jun 17;47(8):2458–2464. doi: 10.1128/JCM.01863-08

TABLE 6.

Sensitivities and specificities for the three tests

Test % Sensitivitya % Specificityb
Amplicor 96.4 91.7
Proofer 64.1 96.7
Linear Array 98.4 NAc
a

Based on 643 women with histologically confirmed CIN2+.

b

Based on 736 women with normal cytology.

c

NA, not available. Specificity could not be measured using the Linear Array as only some (n = 78 HPV positive) of the 736 women with normal cytology were tested via this technique.