abstract 76. Performance of standard minimally invasive surgery tools and the tactile sensing system in a test model to locating occult tumours during video-assisted thoracoscopic surgery

Tools Measure; mean (SD) Measure; % Positive likelihood ratio p value*
Maximum pressure, N/cm2 Average distance, cm Acc. Sens. Spec.
MIS
Manual N/A 0.33 (0.2) 88 88 89 7.88 < 0.01
Grasper 7.2 (2.8) 0.47 (0.3) 51 67 29 0.94 0.79
Ultrasound 1.6 (0.8) 0.68 (0.4) 58 67 42 1.14 0.62
TSS
Liver 2.1 (0.8) 0.46 (0.2) 61 81 29 1.14 0.29
Lung 2.5 (1.0) 0.64 (0.4)§ 57 66 33 0.98 0.95

Acc. = accuracy; MIS = minimally invasive surgery; Sens. = sensitivity; Spec. = specificity; TSS = tactile sensing system.

*

Pearson χ2 (2-sided).

p < 0.01 determined by one-way analysis of variance.

p = 0.13 determined by an independent samples t test (unequal variances) when compared with TSS using liver.

§

p = 0.06 determined as above.