. 2000 Feb 5;320(7231):347–348. doi: 10.1136/bmj.320.7231.347

Table.

Test performance calculated according to different reading techniques (at 450 nm or double reading at 450/655 nm or by direct visual reading without use of plate reader) and to different cut off values as suggested by manufacturer

Reading technique Cut off according to manufacturer Result
Sensitivity (%) (95% CI) Specificity (%) (95% CI) Accuracy (%) (95% CI) Likelihood ratio (95% CI)
No of cases in grey zone*
True positive False negative True negative False positive
Positive Negative
Reading at 450 nm Negative <0.140, grey zone 0.140-0.160, positive >0.160 52 0 136 10 100 (93.2 to 100) 93 (87.8 to 96.7) 95 (90.9 to 97.6) 15 (7.6 to 30.3) 0 (0 to 0.08) H pylori positive=0H pylori negative=4
Reading at 450/655 nm Negative <0.100, grey zone 0.100-0.120, positive >0.120 52 0 136 10 100 (93.2 to 100) 93 (87.8 to 96.7) 95 (90.9 to 97.6) 15 (7.6 to 30.3) 0 (0 to 0.08) H pylori positive=0H pylori negative=4
Direct visual reading NA 49 3 144 2 94 (84.1 to 98.8) 99 (95.1 to 99.8) 97 (94.2 to 99.2) 69 (17.2 to 494) 0.06 (0.01 to 0.17) NA

NA=Not applicable. 

*

According to manufacturer, these cases should be considered equivocal and second stool sample should be tested. The four H pylori negative results are included in the10 false positives. 

According to invasive tests (histology plus urease test or culture).