Skip to main content
. Author manuscript; available in PMC: 2010 Sep 1.
Published in final edited form as: Cytometry B Clin Cytom. 2009 Sep;76(5):295–314. doi: 10.1002/cyto.b.20480

Figure 2. 7-Laser optical bench configuration.

Figure 2

This experimental design platform permits multiple air-cooled lasers to be placed within a modestly sized laboratory benchtop configuration. Each of the lasers has independent steering optics that permits each beam to be accurately focused into the instrument’s flow cell [lower right]. The power and wavelength of each laser, and the number of PMTs utilized is shown on the accompanying table.

Laser Wavelength [nm] Style Power [mw] PMT Array PMT#
1 UV 355 Mode Lock 20 Trigon 2
2 Violet 405 Diode Pumped Solid State 50 Octagon 8
3 Blue 488 Diode Pumped Solid State 20–100 Octagon 4
4 Green 532 Diode Pumped Solid State 40–150 Octagon 5
5 Yellow 594 Diode Pumped Solid State 50 Trigon 2
6 Red 638 Diode Pumped Solid State 40 Trigon 3
7 IR 785 Diode Pumped Solid State 25 Trigon 1