TABLE 2.

Troubleshooting table.

Problem Possible reason Solution
No high molecular weight proteins after
cell lysis
Protein degradation Make sure protease inhibitors were added before
lysis. Avoid sample heating during lysis. Perform
lysis steps up to protein digestion (Steps 1–7) the
same day and as fast as possible
No peptides observed in checkpoint 2
(Box 1)
Protein digestion failed Check integrity/reactivity of trypsin solutions.
pH during digestions should be 7.5–8.5 and urea
concentration <2 M
Peptides elute too late when analyzed
by LC-MS/MS
Maximum capacity was reached on the
SepPak cartridge used for peptide
desalting
Use higher-capacity SepPak cartridge or save the
flow-through and washes and perform a second
desalting step with them
Peptides elute too early in the SCX.
Peptides with high solution charge are
observed in the early fractions
SCX column has not been equilibrated
properly
Equilibrate the SCX column for a longer time
SCX buffers A and B have pH > 3 Make new buffers and ensure they have been
properly prepared
SCX buffer A has higher concentration of
cations
Many nonphosphorylated peptides (>50%)
are detected after IMAC enrichment
Too much IMAC resin was used for the
amount of peptides
Reduce the amount of resin
Incubation time of peptides with IMAC
resin was too long
Do not incubate for longer than 60 min
IMAC resin was not properly washed
before elution
Add an extra wash step with IMAC-binding buffer
(i.e., repeat Step 34A(ii) or 34B(ii))
Very few total peptides were detected Starting amounts of protein were
not sufficient
Although lower amounts might work, we recommend
starting with at least 5 mg of protein
IMAC resin is too old PhosSelect IMAC resin shelf life is less than 1 year.
Store in small aliquots at −20 °C and do not freeze and
thaw multiple times. When performance is decreased,
replace it with a new batch (an SCX fraction prepared
following this protocol can be used as a control)
There was a problem in the LC-MS/MS
system
Make sure LC-MS/MS system is working properly with a
complex mixture standard (see EQUIPMENT SETUP)