TABLE 2.
Problem | Possible reason | Solution |
---|---|---|
No high molecular weight proteins after cell lysis |
Protein degradation | Make sure protease inhibitors were added before lysis. Avoid sample heating during lysis. Perform lysis steps up to protein digestion (Steps 1–7) the same day and as fast as possible |
No peptides observed in checkpoint 2 (Box 1) |
Protein digestion failed | Check integrity/reactivity of trypsin solutions. pH during digestions should be 7.5–8.5 and urea concentration <2 M |
Peptides elute too late when analyzed by LC-MS/MS |
Maximum capacity was reached on the SepPak cartridge used for peptide desalting |
Use higher-capacity SepPak cartridge or save the flow-through and washes and perform a second desalting step with them |
Peptides elute too early in the SCX. Peptides with high solution charge are observed in the early fractions |
SCX column has not been equilibrated properly |
Equilibrate the SCX column for a longer time |
SCX buffers A and B have pH > 3 | Make new buffers and ensure they have been properly prepared |
|
SCX buffer A has higher concentration of cations |
||
Many nonphosphorylated peptides (>50%) are detected after IMAC enrichment |
Too much IMAC resin was used for the amount of peptides |
Reduce the amount of resin |
Incubation time of peptides with IMAC resin was too long |
Do not incubate for longer than 60 min | |
IMAC resin was not properly washed before elution |
Add an extra wash step with IMAC-binding buffer (i.e., repeat Step 34A(ii) or 34B(ii)) |
|
Very few total peptides were detected | Starting amounts of protein were not sufficient |
Although lower amounts might work, we recommend starting with at least 5 mg of protein |
IMAC resin is too old | PhosSelect IMAC resin shelf life is less than 1 year. Store in small aliquots at −20 °C and do not freeze and thaw multiple times. When performance is decreased, replace it with a new batch (an SCX fraction prepared following this protocol can be used as a control) |
|
There was a problem in the LC-MS/MS system |
Make sure LC-MS/MS system is working properly with a complex mixture standard (see EQUIPMENT SETUP) |