Skip to main content
. Author manuscript; available in PMC: 2010 Jun 1.
Published in final edited form as: Magn Reson Med. 2009 Jun;61(6):1500–1506. doi: 10.1002/mrm.21800

Figure 3.

Figure 3

Min-max shim result vs least-squares shim result. When the TR was 10 ms, the bSSFP images from the least-squares shim show the banding artifact (arrows) near the sinus area (a). The artifact disappeared when the shim was reconfigured to the min-max shim for the same TR (b). In the least-squares shim, the TR needed to be shortened to 7 ms to yield artifact free images (c).