Published in final edited form as: Magn Reson Med. 2009 Jun;61(6):1500–1506. doi: 10.1002/mrm.21800

Table 1.

Summary of the results

Experiment 1 (TR = 5 to 10 ms) Experiment 2 (TR = 5 ms)
Least-squares Shim method Min-max Shim method Least-squares shim method Modified min-max shim
Off-resonance (Hz) TRa,b (ms) Off-resonance (Hz) TRa,b (ms) cOff-resonance (Hz) cOff-resonance (Hz)
1 −42.0 to 49.7 8(< 5) −39.5 to 36.4 9(< 5) −62.4 to 113.2 −86.8 to 76.5
2 −45.2 to 58.5 6(< 5) −48.4 to 48.6 7(5) −48.2 to 135.7 −81.1 to 79.1
3 −42.7 to 78.0 5(5) −43.6 to 43.7 8(6) −62.9 to 112.0 −76.0 to 76.0d
4 −33.6 to 54.5 7(7) −36.6 to 36.2 10(9) −51.8 to 127.8 −83.7 to 78.2
5 −45.7 to 62.7 7(< 5) −50.1 to 52.3 9(6) −53.2 to 115.1 −81.1 to 77.4
a

The maximum TR that showed no significant banding artifact in the bSSFP imaging.

b

The number in the parenthesis represents a TR that shows no discernable banding artifact even at the edges of the images.

c

Off-resonance frequency ranges are based on the ROI after removing excessive off-resonant voxels by the modified min-max shim method.

d

The off-resonance frequency range was calculated from the field map.