Table 2.

Repeat HPV-QUAD Results for Specimens Reported as Negative (<2.0) but with Intermediate Signal Ratios (>1.0 <2.0)

Specimens with positive repeat results
Run 2
Run 3
Specimen # HPV type(s) Run 1* Ratio Ratio Result Ratio Result
1 52/58 1.3 2.6 pos
2 31 1.0 10.4 pos
3 35/68 1.4 3.3 pos
4 35/68 1.6 2.2 pos
5 6/11 1.9 2.5 pos
6 35/68 1.6 2.2 pos
7 52/58 1.7 4.5 pos
8 52/58 1.6 2.6 pos
9 31 1.5 4.4 pos
10 39/56 1.9 1.9 neg
51/59 1.2 2.1 pos
52/58 1.7 2.5 pos
11 18 0.0 8.6 pos 0.0 neg
51/59 1.5 7.2 pos 3.1 pos
12 16 0.4 3.4 pos
39/56 1.1 11.6 pos
13 16 1.0 4.1 pos
18 1.8 0.0 neg
39/56 1.2 3.4 pos
Specimens with negative repeat results
Run 2
Specimen # HPV type(s) Run 1 Ratio Ratio Result
14 35/68 1.1 1.0 neg
15 51/59 1.3 1.7 neg
16 39/56 1.4 0.4 neg
17 6/11 1.4 1.7 neg
18 18 1.6 1.9 neg/IND

All specimens were positive in the HC2 assay, except Specimen 13.

*

All Specimens in Run 1 were reported as negative with HPV ratios <2.0.

The result for Specimen 11on Run 2 indicated the presence of HPV18. This was likely contamination and was not present in Run 3.

Specimen 18 was indeterminate (IND) on Run 2 (insufficient beta globin amplification).