Published in final edited form as: Anal Chem. 2009 Jul 1;81(13):5421–5425. doi: 10.1021/ac900732w

Table 1.

Accuracy and Precision of the LC–APCI/MS Method, Based on Recovery of Known Amounts of Phylloquinone Added to Plasma Samples (n = 3)

unlabeled phylloquinone
deuterium-labeled phylloquinone
spiked concentration (nmol/L) mean recovery (%)a RSD (%)b mean recovery (%) RSD (%)
0.5 97.8 ± 15.1 15.4 92.2 ± 7.2 7.8
1.0 92.4 ± 3.2 3.5 87.7 ± 7.7 8.8
2.0 98.8 ±4.3 4.4 96.0 ± 7.7 7.9
4.0 97.7 ± 3.1 3.1 93.5 ± 9.0 9.6
8.0 91.2 ± 1.2 1.4 99.6 ± 8.8 8.7
16.0 95.9 ± 7.1 7.3 97.6 ± 2.6 2.7
32.0 103.2 ± 2.8 2.7 102.0 ± 3.4 3.3
a

Mean percent relative recovery (mean % ± SD) represents the accuracy of the method.

b

Percent relative standard deviation (RSD) represents the precision of the method.