Skip to main content
. 2009 Sep;4(9):1441–1448. doi: 10.2215/CJN.02790409

Table 6.

Small solute clearances in uremic pigs WAK V1.0 vs. WAK V1.1

Results V 1.0 (n = 12) V 1.1 (n = 6) Units P value
Effective urea clearance 24.5 ± 9.8 37.0 ± 7.3 [ml/min] 0.0140
Effective creatinine clearance 18.5 ± 3.2 27.0 ± 4.0 [ml/min] 0.0001
Effective potassium clearance 32.6 ± 7.1 50.3 ± 14.0 [ml/min] 0.0024
Effective phosphate clearance 13.9 ± 1.8 23.1a [ml/min] N/A
Effective urea removal 22.0 ± 8.2 31.3 ± 7.9 [mg/min] 0.0357
Effective creatinine removal 1.5 ± 0.2 2.4 ± 0.3 [mg/min] 0.0005
Effective phosphate removal 1.4 ± 0.2 2.5a [mg/min] N/A
Effective potassium removal 1.9 ± 0.4 11.0 ± 2.9 [mg/min] <0.0001
Hourly Kt/V 0.040 ± 0.004 0.045 ± 0.002 N/A N/A

8 h of WAK dialysis, in anesthetized uremic pigs. WAK, wearable artificial kidney.

a

Data available for one pig only, with average concentration in/out 11.01/5.83 mg/dl, average hematocrit 33%, average blood.