Table 6.
Results | V 1.0 (n = 12) | V 1.1 (n = 6) | Units | P value |
---|---|---|---|---|
Effective urea clearance | 24.5 ± 9.8 | 37.0 ± 7.3 | [ml/min] | 0.0140 |
Effective creatinine clearance | 18.5 ± 3.2 | 27.0 ± 4.0 | [ml/min] | 0.0001 |
Effective potassium clearance | 32.6 ± 7.1 | 50.3 ± 14.0 | [ml/min] | 0.0024 |
Effective phosphate clearance | 13.9 ± 1.8 | 23.1a | [ml/min] | N/A |
Effective urea removal | 22.0 ± 8.2 | 31.3 ± 7.9 | [mg/min] | 0.0357 |
Effective creatinine removal | 1.5 ± 0.2 | 2.4 ± 0.3 | [mg/min] | 0.0005 |
Effective phosphate removal | 1.4 ± 0.2 | 2.5a | [mg/min] | N/A |
Effective potassium removal | 1.9 ± 0.4 | 11.0 ± 2.9 | [mg/min] | <0.0001 |
Hourly Kt/V | 0.040 ± 0.004 | 0.045 ± 0.002 | N/A | N/A |
8 h of WAK dialysis, in anesthetized uremic pigs. WAK, wearable artificial kidney.
Data available for one pig only, with average concentration in/out 11.01/5.83 mg/dl, average hematocrit 33%, average blood.