Table 2.

ENNPS correlation with tomosynthesis scan parameters. The symbol “+” or “−” indicates an increase or decrease in the ENNPS as the scan parameter is increased (i.e., positive or negative correlation, respectively). An “X” is used to indicate a table entry that does not apply to a given imaging technique.

  N ANG NP Plane location (distance from detector)
Raw projections None X X
Conventional tomosynthesis + at low frequencies, − at mid and high frequencies − at low frequencies, + at mid and high frequencies X Increased low-frequency noise in anterior planes, and mid- and high-frequency noise spikes in central and posterior planes
MITS − at low and mid frequencies + at low and mid frequencies − at low and mid frequencies Increased low- and mid-frequency noises in both anterior and posterior planes