Table 2.
ENNPS correlation with tomosynthesis scan parameters. The symbol “+” or “−” indicates an increase or decrease in the ENNPS as the scan parameter is increased (i.e., positive or negative correlation, respectively). An “X” is used to indicate a table entry that does not apply to a given imaging technique.
N | ANG | NP | Plane location (distance from detector) | |
---|---|---|---|---|
Raw projections | None | − | X | X |
Conventional tomosynthesis | + at low frequencies, − at mid and high frequencies | − at low frequencies, + at mid and high frequencies | X | Increased low-frequency noise in anterior planes, and mid- and high-frequency noise spikes in central and posterior planes |
MITS | − at low and mid frequencies | + at low and mid frequencies | − at low and mid frequencies | Increased low- and mid-frequency noises in both anterior and posterior planes |