. 2009 Feb 12;36(3):753–764. doi: 10.1118/1.3075819

Table 4.

Accuracies (root-mean squared error) determined through simulation over 20 000 independent trials in a water∕fat signal at a water SNR of 20. The accuracy of chemical shift is within 0.01 ppm and that of amplitude is within 1% RMSE for ETLs of 4–16. At ETL≤6, the RMSE for both T2* values is high, especially for the longer T2* peak (water).

No. of Echoes Water PRF (ppm) Water T2* (% error) Water amplitude (% error) Fat PRF (ppm) Fat T2* (% error) Fat amplitude (% error)
4 8.28×10−3 42.1 0.425 7.70×10−3 16.6 0.932
5 6.74×10−3 38.4 0.322 6.60×10−3 0.426 0.785
6 4.69×10−3 6.52 0.227 4.7×10−3 0.347 0.478
7 2.71×10−3 0.474 0.136 2.7×10−3 0.302 0.281
8 2.1×10−3 0.223 0.106 2.1×10−3 0.263 0.219
12 4.70×10−4 0.113 3.95×10−2 4.84×10−4 0.166 8.87×10−2