Skip to main content
. 2009 Jul 15;47(9):3011–3013. doi: 10.1128/JCM.00805-09

TABLE 1.

Comparison of sensitivities of HIV RT assay versions 2 and 3, using plasma samples (n = 242) from HIV-positive individuals

HIV RNA level (copies/ml) No. of samples tested % of samples detected by HIV RT assay
Version 2 Version 3
50-400 64 17 59
401-2,000 57 56 86
2,001-10,000 50 90 96
10,001-50,000 48 100 100
>50,000 23 100 100
>200 199 78 92
>400 178 83 94
>1,000 145 92 98