TABLE 1.

Comparison of Phoenix and Vitek 2 results to reference BMD oxacillin MICs

System testeda % Categorical agreement No. of resistant isolatesb No. (%) VM errorsc No. of susceptible isolatesb No. (%) major errorsd % Essential agreemente No. of instrument oxacillin MIC results within the following log2 dilution of reference method MIC
>−2 −2 −1 Same +1 +2 >+2
Phoenix 99.8 448 1 (0.2) 172 0 (0) 91.6 5 3 10 491 67 42 2
Vitek 2 99.7 448 1 (0.2) 172 1 (0.6) 94.7 6 2 10 511 66 25 0
a

A total of 620 isolates were tested with each system.

b

As determined by the reference BMD method.

c

A VM error was resistance by the reference method but susceptibility by the automated method; percentages are based on the number of resistant isolates.

d

A major error is susceptibility by the reference method but resistance by the automated method; percentages are based on the number of susceptible isolates.

e

Essential agreement is an MIC determined with the Phoenix or Vitek 2 system that is ±1 log2 dilution of the MIC determined by the reference method.