Skip to main content
. Author manuscript; available in PMC: 2009 Sep 11.
Published in final edited form as: Med Phys. 2007 Aug;34(8):3334–3344. doi: 10.1118/1.2756612

Table III.

Estimation of the statistical significance in the difference between the FROC performance of the unilateral and bilateral CAD systems on test subsets 1 and 2. The FP rates of the FROC curves were estimated from the no-mass data set: (a) all cases and (b) malignant cases.

(a) FOM (JAFROC)
Test subset 1 Test subset 2
Unilateral CAD 0.52 0.48
Bilateral CAD 0.58 0.51
p value <0.001 0.008
(b) FOM (JAFROC)
Test subset 1 (malignant only) Test subset 2 (malignant only)

Unilateral CAD 0.56 0.53
Bilateral CAD 0.61 0.56
p value 0.009 0.003