Table 1.

Results of the Standard NEMA measurements and Head Equivalent measurements

Fluoroscopy DA DSA
Std Head Eq. Std Head Eq Std Head Eq.
Static features Bar pattern 2.8 3.1 3.1 3.1 3.1 3.1
Iodine C-d target 1 6/8 8/8 8/8 8/8 8/8 8/8
Iodine C-d target 2 6/8 6/8 6/8 6/8 8/8 8/8
Iodine C-d target 3 0/8 0/8 0/8 0/8 4/8 2/8
Iodine C-d target 4 0/8 0/8 0/8 0/8 0/8 0/8
Air cylinders + air Pins 2/4 3/4 2/4 3/4 1/4 0/4
Al cylinders + Pb Pins 4/4 4/4 4/4 4/4 4/4 4/4

X-ray parameters kVp 80 80 73 74 80 96
Tube Current (mA) 32 37 269 342 160 125
Frame Exposure (ms) 9.3 7.4 5.1 4.1 71.1 50
EEphantom (mR/frame) .988 0.760 14.12 10.28 97.2 74.4
EEDetector (mR/frame) 0.011 0.033 0.136 0.231 1.15 2.28

Moving features Moving wires 2/5 4/5 4/5 5/5 4/5 5/5
Temporal Filtering 4 4 None None None None

X-ray parameters kVp 80 80 73 80 93 94
Tube Current (mA) 28 29 231 200 320 320
Frame Exposure (ms) 5.6 4.5 4.1 5.9 16 14.3
EEphantom (mR/frame) .435 0.615 7.88 7.47 45 60
EEDetector (mR/frame) .012 0.028 0.098 0.191 1.456 2.24