Table 3.

Linewidthsa (mG) of signals for trityl-CD3 and LiPc calculated from spectral slices

Trityl-CD3 LiPc
Datab 5,000 50,000 5,000 50,000
30 60 30 60 30 60 30 60
FBP c 36 35 36 35 51 56 53 53
CMEM c 35 35 35 34 56 52 54 52
CMEM2 c 36 34 35 34 53 45 53 50
SMEM c 34 34 34 34 51 50 49 50
AUX c 36 34 36 34 53 55 56 53
Direct d 35 35 35 35 54 54 54 54
a

Linewidths are reported as the full width at half height of the absorption signal. There are 200 data points per spectral slice of 0.34 G. By interpolation it was possible to calculate linewidths with an uncertainty of about 1 mG.

b

30 and 60 are the numbers of projections. 5,000 and 50,000 are the numbers of averages.

c

Linewidths calculated from spectral slices taken images reconstructed by the methods defined in the text.

d

Measured directly from EPR spectra for LiPc and trityl.