Table 1. Effect of shock vector and can position on predicted DFTs and myocardial voltage gradients using standard, two-coil transvenous electrode.
Electrode configuration | Joules | % myocardium >30V/cm |
---|---|---|
RIC → RV + SVC | 93.2 | 18.4 |
RIC + SVC → RV | 14.3 | 30.3 |
RIC + RV → SVC | 22.9 | 32.2 |
LIC → RV + SVC | 15.1 | 4.9 |
LIC + SVC → RV *** | 8.3 | 18.3 |
LIC + RV → SVC | 40.3 | 44.4 |
RAC + RV → SVC | 10.8 | 19.7 |
RAC → RV + SVC | 13.4 | 4.0 |
RAC + SVC → RV | 11.9 | 28.6 |
LAC + RV → SVC | 10.0 | 27.1 |
LAC + RV → SVC | 7.5 | 14.3 |
LAC → RV + SVC | 11.0 | 3.6 |
Can Position Key
LIC: Left infraclavicular
RIC: Right infraclavicular
LAC: Left abdominal
RAC: Right abdominal
Electrode Type Key
RV: 5cm right ventricular electrode
SVC: 8 cm superior vena cava electrode
Standard configuration