Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2009 Aug 18;877(27):3169–3174. doi: 10.1016/j.jchromb.2009.08.011

Table 2. Intra-day and inter-day accuracy and precision of LTE4 assay using calibration standards or pooled spiked urine from asthmatic children.

Intra- and inter-day analyses were conducted over a six-day period using either calibration standards or spiked, pooled urine from asthmatic children. Calibration spike standards were prepared in 30:70 Water:Methanol in 0.08% acetic acid and 0.04% ammonium hydroxide. Standards were then spiked into HPLC water and pooled urine from asthmatics at the indicated concentrations before analysis. Individual results were averaged and coefficient of variation (CV) and percent recovery (%R) were determined. Abbreviations: Average (AVG), Percent recovery (%R).

Calibration standards Intra-day Accuracy and Precision (n=6) Inter-day Accuracy and Precision (n=6)
(pg/mL) AVG CV %R AVG CV %R
5 5.35 15.90 107.07 5.25 11.52 104.92
10 10.11 9.82 101.07 9.77 11.28 97.74
25 24.13 2.59 96.52 24.77 11.69 99.09
50 49.26 7.01 98.52 50.32 5.84 100.64
100 98.66 1.69 98.66 99.35 4.43 99.35
250 238.32 4.00 95.33 244.26 2.18 97.71
500 514.17 3.11 102.83 506.66 1.38 101.33
Spiked urine Intra-day Accuracy and Precision (n=6) Inter-day Accuracy and Precision (n=6)
(pg/mL) AVG CV %R AVG CV %R
+0 81.09 4.59 NA 82.62 10.20 NA
+25 111.31 6.84 120.87 106.16 9.53 94.15
+50 132.11 6.08 102.04 131.67 5.30 98.09
+100 184.98 4.34 103.90 180.31 4.52 97.69
+250 322.11 6.72 96.41 317.71 8.43 94.04
+500 571.85 2.17 98.15 584.47 7.31 100.37