Table 1.

Comparison of flow velocities measured by PC-MR and timed collection through 4mm tube.

Collected Flow Measurement mean ± SD (mL/s | cm/s) PC-MR Flow Calculation mean ± SD (mL/s | cm/s) RMS Error (%)
1.97 ± 0.02 15.7 ± 0.2 (n=14) 1.89 ± 0.23 15.0 ± 1.8 12%
2.23 ± 0.02 17.7 ± 0.2 (n=8) 2.30 ± 0.43 18.3 ± 3.4 18%
2.48 ± 0.01 19.7 ± 0.1 (n=14) 2.36 ± 0.39 18.8 ± 3.1 16%
15%