Skip to main content
. Author manuscript; available in PMC: 2009 Oct 12.
Published in final edited form as: Anal Chem. 2009 Apr 1;81(7):2760–2767. doi: 10.1021/ac802707u

Figure 7.

Figure 7

Sensor output at varying ALC concentrations was quantified using fluorescence microscopy at the microchannel detection port. Normalized averages ±SE were plotted and fit to a nonlinear equation representing combined kinetic and surface adsorption concepts. The solid line under the curve represents the control (+1SE, dotted line). The sensor was able to detect down to 20 pg ALC per milliliter of buffered solution (all points, P < 0.05). Inset: 20 pg/mL−5 µg/mL, linear scale.