Skip to main content
. Author manuscript; available in PMC: 2010 Aug 1.
Published in final edited form as: J Clin Neurophysiol. 2009 Aug;26(4):218–226. doi: 10.1097/WNP.0b013e3181b2f29d

Table 1.

Performance of the system for seizure subjects.

Subjects True
detections
Missed
Seizures
False
detections
%
Sensitivity
Data
duration
(hr)
False
detection
rate (/ hr)
1 10 0 0 100 2.01 0
2 9 1 2 90 1.5 1.33
3 9 1 1 90 0.78 1.28
4 8 1 5 89 0.9 5.56
5 2 0 1 100 0.75 1.33
6 6 0 0 100 1.67 0
7 10 0 2 100 1.75 1.14
8 6 4 1 60 1 1
9 2 3 1 40 1.83 0.55
10 5 0 2 100 1.5 1.33
11 8 2 1 80 0.95 1.05
12 11 0 0 100 1.43 0
13 9 1 2 90 1 2
14 10 0 1 100 1.33 0.75
15 6 0 0 100 2.03 0
16 1 0 0 100 2 0
17 1 0 1 100 1 1
18 9 1 2 90 1.03 1.94
19 0 1 0 0 0.82 0
20 5 5 0 50 0.7 0
21 7 3 1 70 0.9 1.11
22 3 7 2 30 0.94 2.13
23 1 0 0 100 0.75 0
24 8 2 1 80 1.65 0.61
25 1 0 0 100 0.17 0
26 6 3 0 66.7 1.77 0
27 3 5 0 37.5 0.93 0
28 7 3 0 70 0.5 0
Total 163 43 26 79.8 (av.) 33.6 0.86 (av.)