Skip to main content
. Author manuscript; available in PMC: 2010 Oct 1.
Published in final edited form as: J Vasc Interv Radiol. 2009 Aug 19;20(10):1343–1351. doi: 10.1016/j.jvir.2009.05.038

Figure 3.

Figure 3

The greatest increase in lethal ice percentage for either 1.7 or 2.4 mm cryoprobes is obtained when adding the second cryoprobe. Continued large increase in lethal ice is also seen when the third 1.7 mm cryoprobe is adding, suggesting also their weaker freeze capacity.