Table 2.
Optimized p-type epitaxial piezoresistive cantilevers for several choices of instrumentation amplifier. Optimization was performed for operation from 1 Hz to 50 kHz with a minimum thickness of 1 μmW, and maximum bias voltage of 10 V. Three amplifiers were considered: TI INA103, AD622, and AD623. The amplifiers were chosen to represent several price points and noise levels. Although the integrated noise of the AD622 and AD623 are 3 times and 17 times greater than that of the INA103, the minimum detectable force only increases by 10% and 70%, respectively. As the instrumentation amplifier noise increases, the piezoresistor is made shorter and thinner, which increases the sensitivity, while the increased 1∕f noise due to the reduction in N negligibly affects the overall noise while it is less than the amplifier noise.
Cantilever | 1 | 2 | 3 |
Amplifier | INA103 | AD622 | AD623 |
fmin (Hz) | 1 | 1 | 1 |
fmax (kHz) | 50 | 50 | 50 |
l (μm) | 74 | 74 | 74 |
w (μm) | 2 | 2 | 2 |
t (μm) | 1 | 1 | 1 |
lpr∕l | 0.33 | 0.31 | 0.21 |
tpr∕t | 0.25 | 0.23 | 0.16 |
n (cm−3) | 4.4×1019 | 4.4×1019 | 4.4×1019 |
Fmin (pN) | 134 | 147 | 230 |
SF (V∕N) | 3.2×104 | 3.3×104 | 3.8×104 |
β* | 0.38 | 0.39 | 0.43 |
(V) | 4.3×10−6 | 4.9×10−6 | 8.8×10−6 |
2.9×10−6 | 2.9×10−6 | 2.9×10−6 | |
3.1×10−6 | 3.4×10−6 | 4.9×10−6 | |
4.0×10−7 | 2.0×10−6 | 6.7×10−6 | |
fcorner (kHz) | 5.5 | 6.5 | 13.5 |
N | 5.5×108 | 4.7×108 | 2.2×108 |