Original Plan
Replan
Pt. No. Planned dose (Gy) Measured dose (Gy) Difference (%) Planned dose (Gy) Measured dose (Gy) Difference (%)
1 1.589 1.66 4.47 1.589 1.59 0.06
2 1.448 1.50 3.59 1.446 1.47 1.66
3 1.391 1.46 4.96 1.388 1.41 1.59