Skip to main content
. Author manuscript; available in PMC: 2010 Jul 15.
Published in final edited form as: Int J Radiat Oncol Biol Phys. 2009 Apr 23;74(4):1290–1297. doi: 10.1016/j.ijrobp.2009.02.004

Table 5.

Comparison of reconstructed to measured point dose

Original Plan
Replan
Pt. No. Reconstructed dose (Gy) Measured dose (Gy) Difference (%) Reconstructed dose (Gy) Measured dose (Gy) Difference (%)
1 1.66 1.66 0.00 1.58 1.59 −0.63
2 1.50 1.50 0.00 1.48 1.47 0.68
3 1.45 1.46 −0.68 1.41 1.41 0.00