Table 1.

Glucose Control Measures for Original MPC Version 0.1.5 (Regular Sampling) and Updated MPC Version 1.4.3 (Variable Sampling) of the MPC (N =10; Median Values Are Reported)

Original MPC Version 0.1.5 Updated MPC Version 1.4.3
Mean glucose (mM) 5.55 5.99
Glucose SD (mM) 0.87 1.59
Time in targeta (%) 76 56
Time to target (min) 360 177
Hyperglycemic index (mM) 0.17 0.40
Hypoglycemic episodesb (unitless) 0 2
Subjects with hypoglycemia (unitless) 0 2
a

Target glucose range from 4.4 to 6.1 mM (80 to 110 mg/dl).

b

Blood glucose <2.8 mM (<50 mg/dl).