Skip to main content
. 2008 Sep;2(5):890–895. doi: 10.1177/193229680800200521

Table 3.

Percentage of FSM and HemoCue Measurements Within YSI Measurements (glucose ≥75mg/dl)

System Within 5% Within 10% Within 15% Within 20% (ISO)
FreeStyle 215/419(51.3%) 355/419(84.7%) 403/419(96.2%) 415/419(99.0%)
HemoCue 179/419(42.7%) 321/419(76.6%) 395/419(94.3%) 414/419(98.8%)
Fisher's Exact Test 2-sided p-value .0154 .0038 NS: .2565 NS: > .99
NS, nonsignificant