Skip to main content
. 2008 Sep;2(5):814–818. doi: 10.1177/193229680800200510

Table 4.

Number and Percentage of Meter Results within Zones of the Consensus Error Grid.

Meter System Zone A Zone B Zones C, D & E
OneTouch Vita 821/834 (98.4%) 13/834 (1.6%) 0/834 (0.0%)
OneTouch Ultra2 273/278 (98.2%) 5/278 (1.8%) 0/278 (0.0%)

Zone A = no effect on clinical action (clinically accurate);

Zone B = altered clinical action – little or no effect on clinical outcome;

Zone C = altered clinical action – likely to affect clinical outcome;

Zone D = altered clinical action – could have significant medical effect;

Zone E = altered clinical action – could have dangerous consequences.