Skip to main content
. 2009 Jan;3(1):98–109. doi: 10.1177/193229680900300111

Table 2.

Relative Distribution of PreciSense Glucose Estimates Obtained during Preclinical Experiments Presented in the CG-EGA (Venous Whole Blood Samples Used as a Reference)a

CG-EGA zone Rate-EGA Point EGA


Individual zones Sum zones A+B Individual zones Sum zones A+B
A 61.2% 84.5% 82.9% 99.6%
B 23.3% 16.7%
C 9.5% 0.0%
D 3.1% 0.4%
E 2.9% 0.0%
a

The table summarizes 455 data preclinical trials. (A data set is a PreciSense reading vs a venous blood glucose reading with no correction for physiological time delay between the two different compartments.)