Skip to main content
. 2009 Jan;3(1):98–109. doi: 10.1177/193229680900300111

Table 5.

Mean (Average) RAD and Median RAD from 3-Day Human Trialsa

Glycemic region Mean RAD Median RAD
Hypoglycemic <70 mg/dl 23.7% 22.1%
Euglycemia 15.9% 10.5%
Hyperglycemic >180 mg/dl 15.0% 11.5%
PreciSense All 16.3% 11.4%
a

Results are based on glucose estimates where the calibrations included fixed Δϕmax and KD. Only ϕ0 values was individual for each patient. The table includes 288 correlated data sets obtained during the trials covering all gradients (not limited to ±1 mg/dl/min). (A data set is a PreciSense reading and a venous blood glucose reading with no correction for physiological time delay between the two different compartments).