Flow/Pressure Effect on Geometric Particle Sizes of TI Inhalation Powder Emitted from a MedTone Delivery System

Flow rate N (discharge cycles per1 inhaler) Average 16th percentile size (μm) Average 50th percentile size (μm) Average 84th percentile size (μm)
15.0 liters/min 10 2.47 10.65 27.28
28.3 liters/min 10 1.35 4.20 13.14
40.0 liters/min 10 1.19 3.29 9.59