Skip to main content
. 2009 Sep;3(5):1146–1154. doi: 10.1177/193229680900300519

Table 1.

Sensor Accuracy as Derived by Relative Difference and Absolute Relative Difference from the YSI Overall and for Each Day of Usea

Sensor day of wear Subjects N (%)b Matched pairs n (%)c (%) relative difference (%) absolute relative difference


Mean STD Median Min, max Mean SD Median Min, max
Day 1 27 (39%) 818 (38%) −10.7 25.1 −13.5 −62.4, 186.4 19.3 19.3 15.6 0, 186.4
Day 4 21 (30%) 625 (29%) −4.0 21.6 −6.9 −46.1, 117.5 17.0 14.0 13.7 0, 117.5
Day 7 21 (30%) 705 (33%) −4.9 17.1 −6.4 −44.4, 103.7 13.3 11.8 10.2 0, 103.7
Overall 69 2148c −6.8 21.9 −9.4 −62.4, 186.4 16.7 15.8 13.2 0, 186.4
YSI glucose range Subjects N (%)b Matched pairs n (%)c Mean SD Median Min, max Mean SD Median Min, max
40−80 39 206 5.3 37.8 −1.1 −42.6, 186.4 24.3 29.4 17.8 0.1, 186.4
81−120 57 444 −4.5 24.2 −6.0 −62.4, 131.2 17.3 17.5 13.0 0, 131.2
121−240 69 1154 −9.0 18.4 −10.3 −57.1, 131.3 16.1 12.7 13.0 0, 131.2
241−300 50 339 −11.1 13.6 −12.3 −42.0, 53.1 14.6 9.8 13.2 0, 53.1
301−400 27 175 −10.6 12.6 −10.5 −37.2, 31.8 13.7 9.1 11.5 0.2, 37.2
Overall 69 2318 −7.3 21.6 −9.7 −62.4, 186.4 16.6 15.6 13.2 0, 186.4
a

YSI in-clinic days occurred on days 1, 4, and 7 of the study. Day 1 sensor elapsed time evaluated: 2−30 hours; day 4 sensor elapsed time evaluated: 62−99 hours; and day 7 elapsed time evaluated: 139−168 hours. Glucose measurements are between 40 and 400 mg/dl.

b

N, number of subjects contributing data; n, number of matched pairs.

c

Of the matched pairs, 170 fell outside of day 1, 4, or 7 time windows.