Skip to main content
. 2009 Sep;3(5):1146–1154. doi: 10.1177/193229680900300519

Table 4.

Precision Analysis by Sensor Day of Weara

Sensor wear day No. of matched pairs Percent ARD mean Percent ARD (SD) PCV mean PCV (SD)
Overall 32,088 15.6 15.73 11.03 11.13
Sensor day 1 5,250 15.98 14.93 11.3 10.56
Sensor day 2 5,250 14.35 14.23 10.15 10.06
Sensor day 3 5,125 15.8 15.83 11.17 11.2
Sensor day 4 4,729 16.61 15.04 11.74 10.63
Sensor day 5 4,599 13.77 13.49 9.74 9.54
Sensor day 6 4,018 16.41 18.52 11.6 13.09
Sensor day 7 3,116 16.88 18.83 11.94 13.31
a

One matched pair was outside of the 7-day time window.