. 2009 Sep;3(5):1066–1081. doi: 10.1177/193229680900300510

Table 6.

Comparison of Glycemic Control Performance with Increasing Simulated Blood Glucose Sensor Error for 2 h Measurement Interval

Sensor error Median BG (mg/dl) BG IQR (mg/dl) % measurements within range:

72–126 mg/dl <72 mg/dl <47 mg/dl
±2% 108 97–131 78.3% 3.2% 0.40%
±5% 108 97–131 77.4% 3.3% 0.34%
±7% 108 97–122 76.1% 3.4% 0.45%
±10% 110 95–122 74.6% 3.1% 0.40%
±15% 110 95–126 69.5% 4.8% 0.51%
±20% 110 94–126 65.2% 5.3% 0.40%